Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.
I agree, do not show this message again.ACCELERATOR / AVERAGER CIRCUIT CONTROLS THE RESPONSE TIME OF THERMOPILES
E. Smeu1,*
Affiliation
- University “Politehnica” of Bucharest, Physics Dept., Spl. Indepententei 313, Bucharest 060042, Romania
Abstract
Thermal methods for laser radiometry are today the most used. Unfortunately, they are slow. Fortunately, this draw-back can, in some cases, be overcome electronically. The paper presents a detailed way to do this for thermopiles. The response analysis is performed for two circuit versions. Experimental data are presented for a medium-power thermopile, followed by one of the circuits. The system is part of the first Romanian digital powermeter for lasers (third version)..
Keywords
Response of the thermopile, Response time, PID controller, Circuit tuning.
Submitted at: March 12, 2004
Accepted at: Dec. 6, 2004
Citation
E. Smeu, ACCELERATOR / AVERAGER CIRCUIT CONTROLS THE RESPONSE TIME OF THERMOPILES, Journal of Optoelectronics and Advanced Materials Vol. 6, Iss. 4, pp. 1225-1231 (2004)