Cookies ussage consent
Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.
I agree, do not show this message again.CHARACTERIZATION OF PERMANENT MAGNETS WITH THE PULSED FIELD MAGNETIZER
C. Beatrice1, O. Bottauscio1, E. Ferrara1, F. Fiorillo1,* , E. Patroi2
Affiliation
- Istituto Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
- ICPE-CA Advanced Research Institute for Electrical Engineering, Bucharest, Romania
Abstract
We report on experiments aimed at precise characterization of the hysteresis loop behavior of permanent magnets by combination of different measuring methods. A discussion is made, in particular, on the application of the Vibrating Sample Magnetometer (VSM), the electromagnet-based hysteresisgraph (EMH), and the Pulsed Field Magnetizer (PFM) methods. In this framework, special emphasis is placed on advantages and shortcomings of the PFM method. Magnetic hysteresis loops have been determined on isotropic and anisotropic Ba ferrites and bonded Nd-Fe-B samples using both spherical (VSM and PFM) and cylindrical (electromagnet and PFM) samples. Overall agreement is found between the results obtained with all different measuring approaches and different sample geometries, but for a definite trend of the measured coercive field value on the impressed field rate. This can be very fast in the PFM testing (period T ~ 10 ms) and very slow with VSM testing (T ~ 1800 s), and different magnetic viscosity field contributions are consequently put in evidence. It is shown that PFM can afford both the measurement of the whole sample magnetic moment, exploiting the reciprocity principle, or the conventional fluxmetric measurement in practically sized cylindrical samples. In the latter case, non-homogeneous demagnetizing effects, though accounted for by calculation, can represent a major source of uncertainty when using low aspect ratio test specimens..
Keywords
Magnetic measurements, Permanent magnets, Pulsed field magnetizer.
Submitted at: April 26, 2004
Accepted at: June 22, 2004
Citation
C. Beatrice, O. Bottauscio, E. Ferrara, F. Fiorillo, E. Patroi, CHARACTERIZATION OF PERMANENT MAGNETS WITH THE PULSED FIELD MAGNETIZER, Journal of Optoelectronics and Advanced Materials Vol. 6, Iss. 3, pp. 963-968 (2004)
- Download Fulltext
- Downloads: 2 (from 2 distinct Internet Addresses ).