Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.
I agree, do not show this message again.THERMAL AND INFRARED ANALYSES OF ALUMINOSILICATE GLASS SYSTEMS FOR DENTAL IMPLANTS
A. Gritco1, M. Moldovan2, R. Grecu2, V. Simon1,*
Affiliation
- Babes-Bolyai University, Faculty of Physics, Cluj-Napoca, Romania
- Raluca Ripan Institute of Chemistry, 3400 Cluj-Napoca, Romania
Abstract
Aluminosilicate glasses containing florides are investigated by means of thermal analysis and Fourier transform infrared spectroscopy (FTIR). Both methods evidence different silanisation accomplishment in function of glass composition. The silane adheres better on samples also containing B2O3. The FTIR analysis does not evidence any changes in the vibrational properties of the samples immersed for seven days in simulated body fluid..
Keywords
Dental implants, Aluminosilicate glasses, DTA, FTIR.
Submitted at: Aug. 22, 2005
Accepted at: Nov. 24, 2005
Citation
A. Gritco, M. Moldovan, R. Grecu, V. Simon, THERMAL AND INFRARED ANALYSES OF ALUMINOSILICATE GLASS SYSTEMS FOR DENTAL IMPLANTS, Journal of Optoelectronics and Advanced Materials Vol. 7, Iss. 6, pp. 2845-2847 (2005)
- Download Fulltext
- Downloads: 20 (from 17 distinct Internet Addresses ).