Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.
I agree, do not show this message again.Comparative study of mono and double pulse LIBS configurations analyses for on-site transportable set-up optimization
M. SIMILEANU1, R. RADVAN1,*
Affiliation
- National Institute of Research and Development for Optoelectronics INOE 2000, 409 Atomistilor Str. Magurele, Romania
Abstract
The present paper presents a comparative study on the experimental results collected applying LIBS on three certified reference material samples with different copper concentration. The main interest of the study is the tuning of the LIBS setup for onsite operation, as well as elaboration of documentation for future calibration protocols. The quality of the spectra and the accuracy of the results are evaluated and discussed for different irradiation and detection parameters: the gate width and the gain – from the detection system and the sp (355 nm or 1064 nm) and dp LIBS configuration..
Keywords
On-site LIBS, Copper.
Submitted at: Sept. 3, 2012
Accepted at: Oct. 30, 2012
Citation
M. SIMILEANU, R. RADVAN, Comparative study of mono and double pulse LIBS configurations analyses for on-site transportable set-up optimization, Journal of Optoelectronics and Advanced Materials Vol. 14, Iss. 11-12, pp. 1066-1070 (2012)